Stryker Corporation is recalling 44,099 units of various Disposable Spetzler-Malis Bipolar Forceps due to a labeling error regarding the product's shelf life. While the product labels indicate a 54-month (4.5 year) shelf life, the products are actually only validated for a 36-month (3 year) shelf life. These surgical devices were distributed nationwide and internationally between April 2020 and December 2020.
Using the forceps beyond their validated 3-year shelf life may result in the degradation of the device's sterile barrier or mechanical integrity, potentially leading to surgical site infections or device failure during a procedure.
Contact healthcare provider or manufacturer
If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.
This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.
AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.
Sources: FDA iRES ยท Raw API Response
openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.