Steris Corporation is recalling 32 AMSCO 7052HP and 7053HP Washer/Disinfectors because a wire within the electrical box may shift out of its intended position. This shift can lead to internal electrical arcing (sparking) within the unit. The equipment is used in healthcare facilities for cleaning and disinfecting reusable surgical instruments, utensils, and other medical supplies.
Internal electrical arcing can damage the machine's electrical components and presents a potential fire hazard. While the arcing is contained within the unit and does not pose a direct shock risk to users, it indicates a defect that requires correction to ensure the safe operation of the disinfector.
Intended for cleaning and intermediate-level disinfection of soiled reusable hard-surfaced surgical instruments and utensils.
Intended for cleaning and intermediate-level disinfection of soiled reusable hard-surfaced surgical instruments and utensils.
If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.
This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.
AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.
Sources: FDA iRES ยท Raw API Response
openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.