Smiths Medical PM, Inc. is recalling 16 units of the SurgiVet V6004 NIBP Monitor (also known as the BCI Mini-Torr Plus) due to a manufacturing defect on the internal display board. Specifically, a component called a tantalum capacitor may have been installed with the wrong polarity, which can cause the monitor to suddenly turn off or reset without warning during use. This recall affects monitors distributed in six U.S. states and internationally.
If the monitor suddenly resets or powers down, medical staff may lose critical patient data or monitoring capabilities during a procedure. This failure can result in a delay of treatment or a failure to detect changes in a patient's vital signs.
Technical correction of the display board