Roche Diagnostics Corp. is recalling approximately 86,479 units of the ISE-Electrolyte Systems VetLyte ISE Fluid Pack used in veterinary clinics. These fluid packs, manufactured for IDEXX Laboratories, may have a defective seam in their internal two-chamber pouch that allows chemical reagents to mix prematurely. This defect can prevent the VetLyte electrolyte analyzer from calibrating correctly or, more critically, cause the device to provide inaccurate (biased) test results for animal patients.
The mixing of reagents due to the faulty pouch seam can lead to incorrect electrolyte measurements in animals. This poses a risk of misdiagnosis or improper treatment decisions by veterinary professionals based on biased data.
Manufacturer contact for veterinary device