Medium RiskFDA Device

Pacific Medical Group Inc.: Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocard

Agency Publication Date: February 12, 2020
Share:
Sign in to monitor this recall

Affected Products

Product: Button-Style Tocotransducer (Nautilus), 8-foot Cord 2264HAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Toco Board, NFCM9310; Pacific Medical GE Corometrics Nautilus TOCO Bottom Case, NFCM9315; Pacific Medical GE Corometrics Nautilus TOCO Cable Assembly, NFCM9320; Pacific Medical GE Corometrics Nautilus TOCO Coil with Screws, NFCM9325

All serial numbers serviced between 4/13/15 and 4/2/19.

Product: Loop-Style Tocotransducer (Nautilus) 2264LAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Toco Board, NFCM9310; Pacific Medical GE Corometrics Nautilus TOCO Bottom Case, NFCM9315; Pacific Medical GE Corometrics Nautilus TOCO Cable Assembly, NFCM9320; Pacific Medical GE Corometrics Nautilus TOCO Coil with Screws, NFCM9325

All serial numbers serviced between 1/19/15 and 4/2/19

Product: Button-Style Ultrasound Transducer (Nautilus), 8-foot Cord 5700HAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Ultrasound Board, NFCM9510; Pacific Medical GE Corometrics Nautilus Ultrasound Bottom Case, NFCM9515, Pacific Medical GE Corometrics Nautilus Ultrasound Cable Assembly, NFCM9520; Pacific Medical GE Corometrics Nautilus Ultrasound Crystal, NFCM9535

All serial numbers serviced between 10/20/14 and 4/2/19.

Product: Loop-Style Ultrasound Transducer (Nautilus), 8-foot Cord 5700LAX, that may have been serviced with the following: Pacific Medical GE Corometrics Nautilus Ultrasound Board, NFCM9510; Pacific Medical GE Corometrics Nautilus Ultrasound Bottom Case, NFCM9515, Pacific Medical GE Corometrics Nautilus Ultrasound Cable Assembly, NFCM9520; Pacific Medical GE Corometrics Nautilus Ultrasound Crystal, NFCM9535

All serial numbers serviced between 12/18/14 and 4/2/19.

Product: External Toco Transducer (M1355A), that may have been serviced with the following: Pacific Medical Philips M1355A Toco Bottom Case, NFPH9315; Pacific Medical Philips M1355A Toco Cable Assembly, NFPH9320; Pacific Medical Philips M1355A Toco Gauge, NFPH9340

All serial numbers serviced between 10/20/14 and 4/2/19.

Product: Ultrasound Transducer (M1356A), that may have been serviced with the following: Pacific Medical Philips M1356A Ultrasound Board, NFPH9510; Pacific Medical Philips M1356A Ultrasound Bottom Case, NFPH9515; Pacific Medical Philips M1356A Ultrasound Cable Assembly, NFPH9520; Pacific Medical Philips M1356A Ultrasound M1356A Crystal, NFPH9535

All serial numbers serviced between 12/18/14 and 4/2/19.

Product: Toco M2734A, that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips Avalon Smart M2734A/2735A Toco Gauge Flex, NFPHA9341

All serial numbers serviced between 9/2/15 and 4/2/19.

Product: Toco+ transducer with ECG/IUP capability (wired), that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips Avalon Smart M2734A/2735A Toco Gauge Flex, NFPHA9341

All serial numbers serviced between 9/2/15 and 4/2/19.

Product: Ultrasound Transducer (M2736A), that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips M2736A Avalon Smart Ultrasound Bottom Case, NFPHA9515

All serial numbers serviced between 9/2/15 and 4/2/19.

Product: Toco MP Transducer (M2734B), that may have been serviced with the following: Pacific Medical Philips M2734B Avalon Smart TOCO Bottom Case, NFPHA9315-B

All serial numbers serviced between 9/2/15 and 4/2/19.

Additional Information

Agency: Food and Drug Administration (FDA)
Event ID: 83610
Status: Active
Manufacturer: Pacific Medical Group Inc.
Manufactured In: United States
Units Affected: 10 products (9679; 10911; 6481; 6797; 1603; 1994; 466; 2170; 5897; 1776)

Were You Affected by This Recall?

If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.

Find an Attorney

This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.

AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.

Sources: FDA iRES · Raw API Response

openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.