NIPRO Technical Services, Inc. has recalled approximately 499 units of Conductivity Standard Solution due to a defect that reports incorrect conductivity values. These solutions are used to calibrate meters that measure dialysate used in medical procedures. This recall affects catalog numbers 10001, 10011, 10021, 10032, 10040, 10040c, and 10041. Using an incorrectly calibrated meter could result in reporting inaccurate dialysate conductivity, potentially affecting patient treatment.
If the solution provides incorrect conductivity values during calibration, the meters used to monitor dialysate will report inaccurate readings. This could lead to the preparation of improper dialysis fluid, which may result in serious health complications or injury for patients undergoing dialysis.
Secondary standard solution used for the calibration of conductivity cells together with conductivity meters.
If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.
This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.
AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.
Sources: FDA iRES ยท Raw API Response
openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.