Micro-X Ltd. has recalled 5 units of the Rover Mobile X-ray System (model MXU-RV35). The system's exposure settings may not allow enough time to capture clear, diagnostic-quality images due to hardware variations. Healthcare providers should stop using the affected units and contact the manufacturer to arrange for a system correction or repair.
If the x-ray system fails to produce a usable diagnostic image, a patient's diagnosis could be delayed or they may be subjected to additional, unnecessary radiation exposure from repeat scans. No injuries or incidents have been reported to date.
Firm initiated correction via notification letter.
Units affected: 5
If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.
This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.
AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.
Sources: FDA iRES ยท Raw API Response
openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.