Medline Industries, LP is recalling 60,117 medical procedure convenience kits that contain Stryker's Strykeflow 2 Suction Irrigators. A design change in 2024 allows irrigation solution to leak into the device's handpiece and battery pack during use. This defect can cause the device to leak fluid or emit vaporized saline, which may look like smoke and could lead to device failure or serious injury.
The irrigation solution entering the battery pack and handpiece can cause the device to short-circuit, potentially resulting in burns, electric shock, or the emission of vaporized saline. If the device fails during a surgery, it could cause surgical delays or harm to patients and medical staff.
Recall Z-1085-2026
Recall Z-1086-2026
Recall Z-1087-2026
Recall Z-1088-2026
Recall Z-1089-2026
Recall Z-1090-2026
If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.
This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.
AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.
Sources: FDA iRES ยท Raw API Response
openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.