Medium RiskFDA Device
Medical Devices

Getinge 88-Series Washer-Disinfectors Recalled for Overheating Risk

Agency Publication Date: October 1, 2025
Share:
Sign in to monitor this recall

Summary

Getinge Disinfection Ab has recalled 902 units of its Getinge 88-Series Washer-Disinfector (Model 88-5) because the device may overheat during operation. This risk occurs if the circulation pressure calibration was missed or performed incorrectly and the system's low-pressure alarm fails to trigger. The recall affects Model 88-5 units with electrical heating that are running software version GD 14080 Rev A.

Risk

Incorrect calibration of the circulation pressure can lead to overheating, which may damage the device or surrounding equipment and pose a risk of heat-related hazards during operation.

What You Should Do

  1. This recall affects 902 Getinge 88-Series Washer-Disinfector Model 88-5 units that utilize electrical heating and software version GD 14080 Rev A.
  2. Identify affected units by checking for UDI-DI numbers 07340153700109 or 07340153700116 on the device labeling or within the system settings.
  3. Stop using the recalled device. Contact Getinge Disinfection Ab or your authorized distributor to arrange for a service correction, including the necessary circulation pressure calibration.
  4. Call the FDA at 1-888-463-6332 (1-888-INFO-FDA) for more information regarding this recall.

Your Remedy Options

๐Ÿ“‹Other Action
How to: Stop using the recalled device. Contact the manufacturer or your distributor to arrange return, replacement, or correction.

Affected Products

Product: Getinge 88-Series Washer-Disinfector (Model 88-5)
Variants: Electrical heating version
Model / REF:
88-5
Lot Numbers:
All Lot Numbers with electrical heating
UDI:
07340153700109
07340153700116

Software Version: GD 14080 Rev A. Quantity: 902 units.

Additional Information

Agency: Food and Drug Administration (FDA)
Event ID: 97535
Status: Active
Manufacturer: Getinge Disinfection Ab
Sold By: Hospitals; Surgical Centers; Medical Equipment Distributors
Manufactured In: Sweden
Units Affected: 902 units
Distributed To: Colorado, Massachusetts, New Mexico, New Hampshire
Agency Last Updated: March 10, 2026

Were You Affected by This Recall?

If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.

Find an Attorney

This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.

AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.

Sources: FDA iRES ยท Raw API Response

openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.