Getinge Disinfection Ab has recalled 902 units of its Getinge 88-Series Washer-Disinfector (Model 88-5) because the device may overheat during operation. This risk occurs if the circulation pressure calibration was missed or performed incorrectly and the system's low-pressure alarm fails to trigger. The recall affects Model 88-5 units with electrical heating that are running software version GD 14080 Rev A.
Incorrect calibration of the circulation pressure can lead to overheating, which may damage the device or surrounding equipment and pose a risk of heat-related hazards during operation.
Software Version: GD 14080 Rev A. Quantity: 902 units.
If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.
This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.
AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.
Sources: FDA iRES ยท Raw API Response
openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.