GE HealthCare has initiated a voluntary recall of 479 gantry-based computed tomography (CT) systems, including several models within the Revolution CT and Revolution Apex series. The manufacturer identified a defect that allows blue coolant fluid (glycol) to leak onto the floor near the base of the gantry. While the glycol solution is not corrosive and is not at a harmful temperature, it creates a significant slip-and-fall hazard for hospital staff and patients if the leak is not noticed. No injuries have been specifically reported, but the potential for serious falls exists in a clinical environment.
The coolant fluid (glycol) used in the gantry cooling system can leak and accumulate on the floor. This creates a hidden or slippery surface near heavy medical machinery, which poses a risk of slips, trips, and falls for healthcare providers and patients transitioning onto or off the imaging table.
System ID Numbers: 907729APEX1, 650493REV1, 630266CTRV, 773702REVCT6ED, 312996CTRV2, 1000000018, 617726Y6REV2, 617726Y6REV1, 336890APEX, 704801APEX, 336832APEX, 980993CT1, 910715REV1, 603650APEX1, 603526APEX, 973971REV161, 914787APEX, 718963EDCT, 315473APEX1, 614RMHAPEX1, 724357APEX, 843797REVCT2, 806775APEX, 703776APEX, 414805REVCT, 262577REVCT, 832CT000059, 833CT000155, AH7084CT02, AH1923CT01, BG4549CT04, E416249APEXCTE, 082421140968, 082421140960, 082421140959, DK1044CT12, DK1547CT01, DK1051CT04, CT52261701, M40230104, REV020, CT00486, KRCT54972221, KRCT54893311, 0850211084, NO1068CT04, P00652CT01, PL1672CT03, SA1420CT01, SA2973CT01, CT53972381, A201CT07, AC001CT05, 541010CT02, THCTX24524601, AE1025CT03
Quantity: 250 units. Distributed Worldwide. System IDs include 907729APEX1, 256265STCT, 480342APEX1, 480342APEX3, 949250APEX, 858824REV, 707257APEX1, 714456APEX, 719227APEX1, 970945CT, 203576EDAPEX, 860224EDAPEX, 203276APEX1, 302744APEX, 561737APEX1, 561737APEX2, 786243APEX1, 305323APEX, 904UFNAPEX, 863680APEX, 786596APEX, 813972APEX, 813972APEX2, 813972APEX3, 770219APEX1, 630266CTRV, 773702REVCT6ED, 312996CTRV2, 708245APEX1, 309454APEX1, 1000000018, 1000005353, 815968CTRV, 317621APEX1, 913588APEX1, 913588APEX2, 270444APEXCT, 606451APEX1, 516252APEX, 508971REV, 412607CT1, 248898APEX1, 816271SJCT12, 314251APEX, 336890APEX, 704801APEX, 336832APEX, 336884APEX, 910715REV1, 402559REV6, 603650APEX1, 856886CTAPEX, 856355CTREV, 973971REV161, 973940APEX1, 973831APEX1, 973971APEX3, 505722APEX1, 575556REV1, 702961PCT, 702962RRCT, 718932APEX1, 914787APEX, 718963EDCT, 914737APEX, 631670APEX1, 516MHCTED1, HARLEMAPEXCT, 718226VZAPEX, 315473APEX1, 914681APEXCT1, 330729CTAPX2, 614RMHAPEX1, 614RMHAPEX2, 330841APEX1, 330480APEX1, 580559APEX160, 484526APEX, 717960CTAPEX, 717267CTREV, 724357APEX, 864512APEX1, 864725APEX2, 864560APEX1, 731541APX1, 512544APEX1, 512447APEX, 832325REV, 281477APEX1, 281477APEX3, 713R3ACT49, 713442REVEVO3, 713790APEX2, 713790APEX, 713R3ACT48, 414559REV1, 409772CT, 210757APEX, 806775APEX, 806775APEX1, 713790APEX1, 713442APEX, 713442REVEVO5, 385531CT2, 703504APEX, 434654CT, 703391APEX1, 703776APEX, 804594APEX1, 804267APEX1, 425688REV, 206223REV1, 608890APEXELITE, 262577REVCT, 832CT000234, AUCTX24429383, 832CT000230, AUCTX24566130, 833CT000211, AUCTX24493342, AUCTX24493214, AUCTX24491472, 833CT000155, AH4001CT06, AH1923CT01, 140024CT04, 220826CT05, 110094CT02, 441239APEX1, REV2X2400004CN, REV2X2400076CN, 12906848, BG4549CT04, 604557APEXRT1, 250635APEX, 5349227CT11, 867975APEX, 905522REVCT1, 905522REVCT2, E416249APEXCTE, E416249APEXCT2, E416249APEXCT1, 705522REVCT2, APEX160CMEXZ01, 12044778, DK1039CT02, DK1106CT02, DK1066CT01, DK1009CT01, CT52261701, FI1003CT14, 09974427, 12107537, 08089969, 11525316, 11379817, 259056CT01, HU1766CT01, 12743158, 083021862540324, 7982CT0354, 7982CT0360, 11960464, 08876643, 10025899, 08867431, 09897949, CT00442, CT00494, CT00495, CT00638, CT00468, CT00381, CT00611, CT00637, CT00488, CT00457, CT00516, CT00552, CT00460, CT00486, KRCT54358591, KRCT54358601, 0850211084, KRCT54933611, 0850211108, 0850211121, 0850211094, 110807CT06, 0828210145, LWD0320, LWD0324, 859CT000226, 859CT000190, NO1043CT05, NO1043CT04, 1557CT0318, P00652CT01, 600042CT07, PPR590118, SA2980CT01, SA2973CT01, CT53972381, SGCT54294581, 6073CT0103, SGCTX24467795, SGCTX24467793, 12622329, P002CT03, L007CT02, O009CT06, M007CT03, L001CT01, M001CT04, M001CT02, M004CT02, R002CT12, A201CT07, A201CT05, 222769CT01, 529150CT04, 529150CT02, 850030CT02, THCTX24524601, THCTX24513791, THCTX24476143, 12338981, 10127416, 09850616, 09993031, 10396521, 09904040, 10493318, 09526282, 10127401, 10131630, 10231464, 9830CT0444, VNCTX24476164, VNCT5320632A, VNCTX24515754.
Quantity: 42 units. System IDs include 907729APEX1, 205608CT, 256265STCT, 205699CTREV, 480342APEX, 480342APEX1, 480342APEX3, 11320, 949250APEX, 858657APEX, 858824REV, 858824APEX, 310423CT, 707257APEX1, 714456APEX, 650493REV1, 650493REV2, 818708CTWSO, 310222REVO, 719227APEX1, 970945CT, 203576EDAPEX, 860224EDAPEX, 203276APEX1, 302744APEX, 561737APEX1, 561737APEX2, 786243APEX1, 305323APEX, 904UFNAPEX, 863680APEX, 786596APEX, 786596REV, 813972APEX, 813972APEX2, 813972APEX3, 770219APEX1, 478633REVO, 808522RCT, 630266CTRV, 773702REVCT6ED, 336884APEX.
Quantity: 4 units. System IDs: 907729APEX1, 336832APEX, 973971CTREV, 513584APEX1.
Quantity: 1 unit. System ID: 907729APEX1.
Quantity: 105 units. Distributed Worldwide. System IDs include 907729APEX1, 480342APEX, 480342APEX3, 949250APEX, 858657APEX, 310423CT, 707257APEX1, 650493REV1, 408559CTREVO, 904UFNAPEX, 813972APEX2, 813972APEX3, 808522RCT, 312996CTRV2, 708245APEX1, 718334CTREV, 765298APEX2, 913588APEX2, 270444APEXCT, 508971REV, 413794APCT, 980993CT1, 973971REV160, 702961PCT, 646962APEX1, 516MHCTED1, COR526548, 718226VZAPEX, 330480APEX1, 864725APEX2, 985NODCAPEX, 615936REVO, 713790APEX, 210757APEX, 713442APEX, 703776APEX1, 425688REV, 206320REV2, 206668REV, 608890APEXELITE, 608263REVCT, 414805IPCT, 262577REVCT, 83195527023, AUCTX24493214, AUCTX24491472, 833CT000155, 83155841562, AH1923CT01, 140024CT04, 030355CT07, 190059CT07, 030005CT04, 12906848, 250635APEX, 204726RCT, DK1547CT02, DK1106CT02, DK1009CT01, FI1003CT14, 06405034, C4193357, X143039508, 07272697, 11525316, 11379817, 259056CT01, 11437190, 083021862540324, 7982CT0354, 7982CT0360, 13043139, A5624142, 11960464, 08876643, A5582819, 09897949, A5605766, REV075, CT00494, CT00468, CT00381, REV072, CT00488, CT00516, KRCT54358591, 0850211121, 0850211094, 0850210965, KW1018CT05, 859CT000190, NO1043CT05, NO1043CT04, 6073CT0095, P002CT03, L007CT02, R002CT10, O005CT14, 529150CT04, 529150CT02, 1573CT0126, THCTX24476143, 08124357, 09904040, 10231464.
Quantity: 21 units. System IDs include 907729APEX1, 650493REV2, 864512APEX1, 832325REV, 281MWREVCT, 713792ACT39, 206320REV, 206320REV2, 082421190701, DK1044CT12, DK1547CT02, 11525316, 259056CT01, 11437190, 0850211094, 12622329, M001CT06, M004CT02, AC002CT05, R002CT12, AC001CT05.
If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.
This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.
AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.
Sources: FDA iRES ยท Raw API Response
openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.