Medium RiskFDA Device

GE Healthcare, LLC: When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.

Agency Publication Date: December 13, 2018
Share:
Sign in to monitor this recall

Affected Products

Product: Discovery NM/CT 670 DR, model 5376204-70-56 intended for use in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.

GTIN 00840682124393 Serial Numbers: DDRW34032, DDRX34038, DDRX34041, DDRX34045

Product: Discovery NM/CT 670 ES, model 5376204-70-57

GTIN 00840682123877 Serial Numbers: ESDX35013, ESDW35010

Product: Discovery NM/CT 670 Pro, model 5376204-70-54

GTIN 00840682121194 Serial Number ESDX35015

Additional Information

Agency: Food and Drug Administration (FDA)
Event ID: 81680
Status: Resolved
Manufacturer: GE Healthcare, LLC
Manufactured In: United States
Units Affected: 3 products (4 units; 2 units; 1 unit)

Were You Affected by This Recall?

If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.

Find an Attorney

This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.

AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.

Sources: FDA iRES · Raw API Response

openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.