DRG International, Inc. is recalling 47 units of the T4 Total ELISA test kits (Model EIA-4568) because of a labeling error. The expiration date on the outer kit box and the Certificate of Analysis is incorrectly listed as March 31, 2026, which exceeds the actual shelf life of the test components. The individual bottle labels inside the kit correctly show the expiration date as January 31, 2026. No incidents or injuries have been reported.
Using the test kit after its actual expiration date of January 31, 2026, may result in inaccurate medical test results because the standard and control solutions in the kit could degrade over time.
Expiration date on box and CoA incorrectly listed as 2026-03-31; correct shelf life ends 2026-01-31.
AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.
Sources: FDA iRES ยท Raw API Response
openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.