3M Company has recalled approximately 362,748 units of Attest Super Rapid Readout Biological Indicators and Steam Challenge Packs. The recalled products include specific lots of models 1492V, 1496V, 41482V, and 41482VF. These medical devices are being recalled because the caps on the biological indicators may melt, deform, or crack when subjected to standard steam sterilization cycles at 132°C and 135°C. This defect can compromise the sterilization monitoring process in healthcare settings.
A melted, deformed, or cracked cap can cause the biological indicator to fail or leak during or after the sterilization process. This could lead to inaccurate sterilization monitoring results, potentially resulting in the use of improperly sterilized medical instruments on patients.
Healthcare provider or manufacturer consultation
If you or a family member were harmed by this recalled product, you may have legal rights. Consider consulting a consumer protection attorney to understand your options for compensation.
This is general information, not legal advice. Go Backs is not a law firm and does not provide legal services.
AI-Enhanced Content: The summary, action steps, and risk assessment on this page were generated by AI from official government recall data to improve readability. This is not legal or medical advice. Always refer to the official agency sources below for authoritative information.
Sources: FDA iRES · Raw API Response
openFDA Disclaimer: This recall information is sourced from the openFDA API. Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. Refer to the openFDA terms of service for more information.